Title
In Vitro Assessment of a Breath-synchronized Vibrating Mesh Nebulizer During Non Invasive Ventilation
In Vitro Comparison of Continuous and Breath-synchronized Vibrating Mesh Nebulizer During Non Invasive Ventilation: Analysis of Inhaled and Lost Doses.
Phase
N/ALead Sponsor
Katholieke Universiteit LeuvenStudy Type
InterventionalStatus
Completed No Results PostedIndication/Condition
Respiratory Diseases Cystic Fibrosis COPD Asthma ...Intervention/Treatment
amikacin ...Study Participants
3Using an adult lung bench model of non invasive ventilation, the aim of the study is to compare an experimental system of breath-synchronized vibrating mesh nebulizer to a conventional vibrating mesh nebulizer during non invasive ventilation in terms of inhaled and lost doses.
500 mg/4 mL Amikacin nebulized using vibrating mesh nebulizer associated with a single limb bilevel ventilator. The nebulizations are considered as finished when there is no visible evidence of nebulization for a period of 30 seconds. The NIV is set with an IPAP of 15 cmH2O and EPAP of 5 cmH2O. The lung model is simulated with a respiratory rate of 15 cycles/minute
500 mg/4 mL Amikacin nebulized using vibrating mesh nebulizer associated with a single limb bilevel ventilator. The nebulizations are considered as finished when there is no visible evidence of nebulization for a period of 30 seconds. The NIV is set with an IPAP of 15 cmH2O and EPAP of 5 cmH2O. The lung model is simulated with a respiratory rate of 25 cycles/minute
500 mg/4 mL of Amikacin solution delivered with an experimental breath-actuated vibrating mesh nebulizer associated with a single limb circuit ventilator.
500 mg/4 mL of Amikacin solution delivered with a conventional vibrating mesh nebulizer (in continuous mode) associated with a single limb circuit ventilator.
Inclusion Criteria: Not applicable (in vitro study) Exclusion Criteria: hypersensitivity (allergic) reactions to aminoglycosides